Replicates

If a certain analysis is frequently carried out, it may be advantageous to keep in stock some well-mixed bulk samples (low, medium and high range values), sub-samples of which should be included with every fresh batch of analyses, and again well-mixed before weighing. After a statistically significant number of analyses, an accepted average value for the bulk sample is obtained, and the amount of scatter of results computed as a standard deviation, s, where n = total number of results x = observed value x = mean value of observed concentrations (x - x) = deviation from mean

If subsequent analyses of the bulk sample deviate by more than a predetermined amount, the whole batch of results is rejected. Results are thus only accepted if they fall between specified values of s above and below the mean, where 1s includes 68%, 2s includes 95% (the normally accepted value), and 3s includes 99.7% of results. The scatter of results usually assumes a symmetrical normal or Gaussian distribution about the mean, as shown in Figs 12.1 and 12.2.

Bulk samples are repeated, and if still outside the acceptable limits of precision, the methodology must be examined for sources of error; this was considered fully in an early paper by Büttner (1968).

Scatter of results from replicate analyses with random errors

Concentration

Fig. 12.1. Typical scatter of results from nine replicate analyses. a = absolute error of the determination, d = systematic error, n = total number of results (=9 in Fig. 12.1; see under 'Replicates' in the text), t = true value, x = mean observed concentration value, x = observed value (see under 'Replicates' in text).

about 95%

about 99.8%

Fig. 12.2. A normal or Gaussian distribution of results with % population enclosed by various ± standard deviation values.

With automated segmented flow analysis, the scatter (or distribution) of results often departs from a normal distribution, and may be skewed (Faithfull, 1972). The tendency is for results over a 90-min period (sampling rate 40 h-1) to be negatively skewed, with a tail at lower values and the peak occurring at a higher value. This probably results from changes in the flexing properties of the pump tubing, with 20-40 min of reagent pumping required before an approximately normal distribution of results with an acceptable standard deviation is obtained. Acidflex acid-resistant tubing has been shown to require up to 1 h to stabilize (Davidson et al., 1970). The question of calibration drift and specimen interaction in segmented flow analysis was discussed by Bennet et al. (1970).

Odd sample values occurring some way from the cluster of values around the mean are known as outliers. The problem of whether or not they are acceptable, especially with skewed distributions of data, is considered in AMC (2001), which is also found at:

http://www.rsc.org/lap/rsccom/amc/amc_index.htm Outlier tests, and other statistical methods in analytical chemistry, are also discussed by Meier and Zund (1993).